ɬ﷬

Negative Staining

Optimized Protocol for Negative Staining of Extracellular Vesicles and Particles (EVPs)

Adapted and Prepared by Jeannie Mui

Ultracentrifugation protocol (from Thermo Fisher Scientific)

  1. Collect your EVPs by ultracentrifugation or other means. For ultracentrifugation, spin as needed (e.g., ~120,000 g for 80 to 90 min at 4 °C), pour off supernatant and add conditioned media until the entire volume is processed into a pellet.
  2. Wash the pellet for 30 min at 4°C in 2 to 3 mL of ice-cold 0.1 μm sterile-filtered 1X PBS on a shaker, followed by another spin and one final wash/spin to remove secreted proteins and other components.

Fixation

  1. Resuspend the pellet in 2.5% glutaraldehyde fixative solution in 0.1M sodium cacodylate buffer (final pellet resuspension depends on how much the starting material is, the number of TEM grids you want to use, and the amount of EVPs in your sample; 5 to 10 µL of suspension for each TEM grid. The concentration should be around 1 to 5 μg/μL.
  2. Transfer EVPs to a 1.5 mL Eppendorf test tube and store them at 4 °C.
  3. Bring to the FEMR for negative staining and TEM imaging within one to two days.

Negative staining

1. Grid Preparation

  • Usecarbon-coated 200-mesh Cu TEM grids, carbon side up.
  • Place grids in thePelco easiGlowwith the metal grid holder.
  • Glow dischargefor30 seconds at 30 mAto render the surface hydrophilic.
  • Use gridswithin 20 minutesof glow discharge.

2. Sample Preparation

  • Suggested to prepareat least two grids per specimenwhen possible.
  • If the sample is too concentrated,dilute with PBS or ddH₂Oto a final concentration of0.1 μg/μL.

    ⚠️Note: Avoid phosphate buffers as they cause uranyl acetate precipitation.


3. Sample Application

Place a clean sheet ofparafilmon the glass plate at thenegative staining station (Room SADB B/6).

Choose one of the following methods:

Method 1: Direct Application

  • Using self-locking tweezers, pipette5–10 µLof EVP solution onto thecarbon sideof the grid.
  • Incubate for5 minutes.

Method 2: Drop Incubation

  • Pipette a 20 µLdrop of EVP solution onto parafilm.
  • Place the gridcarbon side downon the drop.
  • Incubate for5–10 minutes, covering with a lid to prevent disturbance.
  • To increase EVP concentration:
    • Incubate for 10 minutes.
    • Remove the grid briefly.
    • Reapply to the same drop for additional accumulation.

4. Washing and Staining

Glycine Wash

  • Transfer the grid (sample side down) ontothree 50 µL drops of 0.2 M glycine,2 minutes each.

Water Wash

  • Transfer tofive 100 µL drops of ddH₂O,1 minute each.

Negative Staining

  • Place the grid (sample side down) on a20 µL drop of filtered 2% uranyl acetatefor1 minute.

    ⚠️Note: Uranyl acetate is acidic; avoid for acid-sensitive particles.

  • Gentlywick off excess stainusing the edge of filter paper.


5. Drying and Storage

  • Allow the grid toair dry for 60 minutes at room temperatureor use aheat lamp.
  • Proceed toTEM imagingor store in agrid boxfor up to1–2 weeks.

Common mistakes or pitfallsthat can occur when following the negative staining protocol for extracellular vesicles and particles (EVPs), along with tips to avoid them:


1. Improper Glow Discharge

  • Mistake:Not using the carbon side up or exceeding the 20-minute window post-discharge.
  • Impact:Leads to poor sample adherence and uneven distribution.
  • Tip:Always confirm the carbon side is facing up and time your sample application promptly.

2. Sample Overload or Underload

  • Mistake:Applying too concentrated or too dilute a sample.
  • Impact:Overload causes clumping; underload results in too few particles for imaging.
  • Tip:Adjust concentration to ~0.1 μg/μL and visually inspect density before staining.

3. Using Phosphate Buffers

  • Mistake:Using PBS or other salt-containing buffers with uranyl acetate.
  • Impact:Causes precipitation and artifacts on the grid.
  • Tip:Use dH₂O for dilution and rinsing before staining.

4. Inconsistent Incubation Times

  • Mistake:Varying incubation times between grids or skipping the reapplication step in Method 2.
  • Impact:Leads to inconsistent EVP density across grids.
  • Tip:Use a timer and standardize incubation steps across all samples.

5. Improper Wicking Technique

  • Mistake:Touching the sample area directly with filter paper.
  • Impact:Can remove or damage the sample.
  • Tip:Wick from theedgeof the grid only.

6. Inadequate Washing

  • Mistake:Skipping or shortening glycine and water washes.
  • Impact:Residual fixatives or salts can interfere with staining and imaging.
  • Tip:Follow the full wash sequence to ensure a clean background and contrast.

7. Inappropriate Use of Uranyl Acetate

  • Mistake:Using uranyl acetate on acid-sensitive particles.
  • Impact:Particle degradation or morphological changes.
  • Tip:Consider alternative stains (e.g., ammonium molybdate) for acid-sensitive samples.

8. Incomplete Drying

  • Mistake:Imaging before the grid is fully dry.
  • Impact:Can cause beam damage or poor image quality.
  • Tip:Ensure at least 60 minutes of drying or use a heat lamp if needed.

Notes

  • Exosomes typically appear cup-shaped due to dehydration and vacuum artifacts during TEM preparation.
  • Ensure all antibody incubations are performed in a humidified environment to prevent drying.
  • Use freshly prepared staining and fixation solutions for optimal results.
Back to top